Universal pyrometer – KTRD 1465 HS

Measuring ranges from 300 °C to 1500 °C

Description

Performance, precision and reliability – you can count on MAURER pyrometers. Benefit from maximum adaptability by choosing the right optics, accessories and the appropriate interface. Thus our pyrometer fits perfectly into existing systems. For tight spaces, this pyrometer is equipped with a fiber optic cable and an attached optic system.

The most important features and characteristics at a glance:

  • Spectral range: 2,1 μm
  • Ultra short response time of 100 μsec
  • Emissivity from 100 – 10 %, adjustable directly at the pyrometer or via interface
  • Green target light (LED) that is exactly identical to the actual measuring spot in position and size
  • With fiber optic cable and optic system for exact focusing even in narrow spaces
  • Use with either digital and analogue output
  • 1 limit output (open collector)
  • Operating voltage 24 V DC
  • Free software (IR-LOG) for parameter setting, data recording and storage

 

Measuring Ranges

1. 150 – 800 °C
2. 200 – 1200°C

Special measuring ranges upon request

 

Scope of Applications
Processing of steel, iron, non-ferrous metal, wires and ceramics; rolling, soldering, welding, transforming, induction heating; vacuum furnace etc.

 

Technical Data

Spectral range 2,1 μm
Target marking Green target light (LED)
Response time 100 μsec
Measurement uncertainty 0,5 % ± 1 °C (ε = 1, Tu = 23°C, T 95 = 1s)
Temperature dependence 0,05 % / °C
Reproducibility 0,1 % ± 1 °C (ε = 1, Tu = 23°C, T 95 = 1s)
Emissivity 100 – 10 %, adjustable directly at the pyrometer or via interface
Resolution < 0,1 % at the analogue output < 0,1 °C at the interface
Operating / Storage temperature 0 – 50 °C, Optics 150 °C / -10 – 70 °C 32 – 140 °F, Optics 302 °F / 14 – 158 °F
Permissible humidity 35 – 85 % RH (non-condensing)
Analogue output 0 – 20 mA / 4 – 20 mA (Load max. 500 Ω)
1 limit output (open collector) 24 V 100 mA
Interface RS 232 ± 50 V isolated or RS 485 ± 70 V isolated
Optional: PROFIBUS, PROFINET, Ethernet, EtherCAT, USB 2.0
Maximum value memory Maximum memory, peak value memory double, adjustable time and threshold value, erasing after time, external contact, by software, after every measuring object
Supply voltage 24 V DC ± 10 %
Power consumption < 100 mA
Device connection 12-pol. plug connector
Dimensions (LxHxW) / Weight 54 x 54 x 147mm / 0,55 kg without fiber optic cable
IP Code IP65

 

Fiber optic cables
Quartz fiber optic cable in metal hose, ambient temperature  max. 150 °C, minimum bending radius = 200 x fiber thickness. Both sides with FSMA-plug for connecting the pyrometer and the optic system. Custom-tailored lengths and other fibers available upon request.

 

Configurations

  • Version with fiber optic cable: KTRD 1465
  • Built-in digital display on the backside of the pyrometer

 

Optic systems

Datasheet

 

Accessory

Mechanical Equipment Electronic Equipment Electronic process units
Blowing device Connection cables AED 1012
Mounting parts (Line-) Scanners AED 1012-C PID Unit
Cooling case Power Supply 100-270 VAC – 24 VDC AED 1012-PC Program-Unit
Mirror 90° for beam deflection PC-Box (USB-connection set)
Optiken