Compact pyrometer – KTRD 4085 HS

Measuring ranges from 600 °C to 1600 °C

Category:

Description

Performance, precision and reliability – you can count on MAURER pyrometers. The compact and easy to use design, as well as a large selection of matching accessories, enables you to quickly integrate the pyrometer into your surroundings.

The most important features and characteristics at a glance:

  • Spectral range:0,85 – 1,1 μm
  • Ultra short response time of 100 μsec
  • Emissivity from 100 – 10 %, adjustable directly at the pyrometer or via interface
  • Green target light (LED) that is exactly identical to the actual measuring spot in position and size
  • With Vario-optic for exact focusing on the measuring spot
  • Use with either digital and analogue output
  • 1 limit output (open collector)
  • Operating voltage 24 V DC
  • Free software (IR-LOG) for parameter setting, data recording and storage

 

Measuring Ranges

Short measuring ranges
1. 600 – 1200 °C (D = 135)
2. 650 – 1300 °C (D = 135)
3. 700 – 1400 °C (D = 135)
4. 750 – 1500 °C (D = 135)

Long measuring ranges
5. 600 – 1600 °C (D = 135)

Special measuring ranges upon request

 

Scope of Applications
Processing of steel, iron, non-ferrous metal, wires and ceramics; rolling, soldering, welding, transforming, induction heating; vacuum furnace etc.

 

Technical Data

Spectral range 0,85 – 1,1 μm
Target marking Green target light (LED)
Response time 125 μsec
Measurement uncertainty 0,5 % ± 1 °C (ε = 1, Tu = 23°C, T 95 = 1s)
Temperature dependence 0,01 % / °C
Reproducibility 0,1 % ± 1 °C (ε = 1, Tu = 23°C, T 95 = 1s)
Emissivity 100 – 10 %, adjustable directly at the pyrometer or via interface
Resolution < 0,1 % at the analogue output < 0,1 °C at the interface
Operating / Storage temperature 0 – 60 °C, Optics 150 °C / -10 – 70 °C 32 – 140 °F, Optics 302 °F / 14 – 158 °F
Permissible humidity 35 – 85 % RH (non-condensing)
Analogue output 0 – 20 mA / 4 – 20 mA (Load max. 500 Ω)
1 limit output (open collector) 24 V 100 mA
Interface RS 232 ± 50 V isolated or RS 485 ± 70 V isolated
Maximum value memory Maximum memory, peak value memory double, adjustable time and threshold value, erasing after time, external contact, by software, after every measuring object
Supply voltage 24 V DC ± 10 %
Power consumption < 100 mA
Device connection 8-pol. plug connector if RS232, 7-pol. plug connector if RS485
Dimensions (LxHxW) / Weight 30 x 65 x 80 mm / 0,19 kg
IP Code IP65

 

Configurations:

  • Version with optic: KTRD 4085
  • Special objectives and measuring ranges upon request

Datasheet

 

Accessory

Mechanical Equipment Electronic Equipment Electronic process units
Blowing device Connection cables AED 1012
Mounting parts (Line-) Scanners AED 1012-C PID Unit
Cooling plate Power Supply 100-270 VAC – 24 VDC AED 1012-PC Programm-Unit
Mirror 90° for beam deflection PC-Box (USB-connection set)
Go to Top