Universal pyrometer – KTRD 1065

Measuring ranges from 100 °C to 1400 °C

Description

Performance, precision and reliability – you can count on MAURER pyrometers. Benefit from maximum adaptability by  choosing the right optics, accessories and the appropriate interface. Thus our pyrometer fits perfectly into existing systems.

The most important features and characteristics at a glance:

  • Spectral range: 2,1 μm
  • Short response time from < 0,5 ms to 5 ms
  • Emissivity from 100 – 10 %, adjustable directly at the pyrometer or via interface
  • Available with either light beam aiming device (LED / Laser) or viewfinder
  • Green target light (LED) that is exactly identical to the actual measuring spot in position and size
  • With Vario-optic for exact focusing on the measuring spot
  • Use with either digital and analogue output
  • 1 limit output (open collector)
  • Operating voltage 24 V DC
  • Free Software (IR-LOG) for parameter setting, data recording and storage

 

Measuring Ranges

1. 100 – 800 °C
2. 150 – 1200 °C
3. 200 – 1400 °C

Special measuring ranges upon request

 

Scope of Applications
Processing of steel, iron, non-ferrous metal, wires and ceramics; rolling, soldering, welding, transforming, induction heating; vacuum furnace etc.

 

Technical Data

Device type KTRD 1065-1 or KTRD 1065-2
Spectral range 2,1 μm
Target marking Light beam aiming device (green LED/Laser) (KTRD 1065-1) or viewfinder (KTRD 1065-2)
Response time Measuring range 1-2: 5ms Measuring range 3: < 0,5ms
Measurement uncertainty 0,5 % ± 1 °C (ε = 1, Tu = 23°C, T 95 = 1s)
Temperature dependence 0,05 % / °C
Reproducibility 0,1 % ± 1 °C (ε = 1, Tu = 23°C, T 95 = 1s)
Emissivity 100 – 10 %, adjustable directly at the pyrometer or via interface
Resolution < 0,1 % at the analogue output < 0,1 °C at the interface
Operating / Storage temperature 0 – 50 °C, Optics 150 °C / -10 – 70 °C 32 – 140 °F, Optics 302 °F / 14 – 158 °F
Permissible humidity 35 – 85 % RH (non-condensing)
Analogue output 0 – 20 mA / 4 – 20 mA (Load max. 500 Ω)
1 limit output (open collector) 24 V 100 mA
Interface RS 232 ± 50 V isolated or RS 485 ± 70 V isolated
Optional: PROFIBUS, PROFINET, Ethernet, EtherCAT, USB 2.0
Maximum value memory Maximum memory, peak value memory double, adjustable time and threshold value, erasing after time, external contact, by software, after every measuring object
Supply voltage 24 V DC ± 10 %
Power consumption < 100 mA
Device connection 12-pol. plug connector
Dimensions (LxHxW) / Weight 54 x 54 x 147mm / 0,6 kg
IP Code IP65

 

Configurations

  • High speed version: KTRD 1065-HS
  • Built-in digital display on the backside of the pyrometer
  • Extensive selection of objectives, suitable for your application

 

Optic systems

Datasheet

 

Accessory

Mechanical Equipment Electronic Equipment Electronic process units
Blowing device Connection cables AED 1012
Mounting parts (Line-) Scanners AED 1012-C PID Unit
Cooling case Power Supply 100-270 VAC – 24 VDC AED 1012-PC Program-Unit
Mirror 90° for beam deflection PC-Box (USB-connection set)
Optiken
Go to Top